    Home » Labkotec SET-2000 110 VAC Supply Voltage User Manual
    Labkotec

    Labkotec SET-2000 110 VAC Supply Voltage User Manual

    Gabriel SmithBy Gabriel SmithNovember 1, 2025No Comments8 Mins Read
    Facebook Twitter Pinterest LinkedIn Tumblr WhatsApp Telegram Email
    Labkotec-SET-2000-110-VAC-Supply-Voltage-featured
    Share
    Facebook Twitter LinkedIn Pinterest Email

    labkotec-logo

    Labkotec SET-2000 110 VAC Supply Voltage

    Labkotec-SET-2000-110-VAC-Supply-Voltage-product

    General Information

    SET-2000 is a one-channel level switch. Typical applications are high-level and low-level alarms in liquid tanks, condensed water alarms, level control, and alarms in oil, sand, and grease separators.

    Labkotec-SET-1000-12-VDC-Level-Switch-Senso-fg-1

    Installation

    The SET-1000 can be wall-mounted. The mounting holes are located in the base plate of the enclosure, beneath the mounting holes of the front cover. The connectors of the external conductors are isolated by separating plates. The plates must not be removed. The plate covering the connectors must be installed after executing cable connections. The cover of the enclosure must be tightened so that the edges touch the base frame. Only then do the push buttons function properly, and the enclosure is tight. Before installation, please read the safety instructions in the chapter!

    Using Junction Box

    This point can be connected to the equipotential ground through the ground terminal. Other components of the system that need to be grounded can also be connected to the same ground terminal. The wire used for equipotential grounding must be a minimum. 2.5 mm²mechanically protected or, when not protected, the minimum cross-section is 4 mm². Please make sure that the cable between SET-1000 and the sensor does not exceed maximum connection values – see Appendix 2. Detailed cabling instructions can be found in the instructions of particular Labkotec SET sensors.

    Labkotec-SET-2000-110-VAC-Supply-Voltage

    Level sensor cabling with a junction box for the cable extension. The junction box of type LJB2 includes light alloy parts.

    Using Cable JointLabkotec-SET-2000-110-VAC-Supply-Voltage

    Please make sure that the sensor and cable between the ET-1000 control unit and the sensor do not exceed the maximum allowed electrical parameters – see Appendix 1, Technical data. IP rating of the cable joint is IP68.

    Labkotec-SET-2000-110-VAC-Supply-Voltage

    Installation accessories, Level sensor cabling with a cable joint for cable extension

    Setting

    See a more detailed description in chapter 3.1 Operation. Channel 1 Alarm takes place when the level hits the sensor (high-level alarm). The trigger level is normally at the middle of the sensor’s sensing element.

    Operation

    If the operation is not as described here, check the settings (chapter 3) and the operation (chapter 4) or contact a representative of the manufacturer. Normal mode – no alarms. The level in the tank is below the sensor.

    The main LED indicator is on. Other LED indicators are off. High-level alarm: The level has hit the sensor (sensor in the medium). The main LED indicator is on. Alarm LED indicator is on. Buzzer on after 5 5-second delay. Relays de-energize after 5 5-second delay.

    Fault alarm: Sensor cable break, short circuit, or a broken sensor, i.,e. too low or too high sensor signal current. The main LED indicator is on. Sensor cable Fault LED indicator is on after 5 5-second delay. The relays de-energize after 5 5-second delay. The buzzer is on after 5 5-second delay. Reset the alarm when pressing the Reset push button. The buzzer will go off. Relay 1 energizes.

    Relay 2 will stay de-energized until the actual alarm or fault is off..The test function provides an artificial alarm, which can be used to test the function of the SET-1000 level switch and the function of other equipment that is connected to SET-1000 via its relays.

    Normal situation

    When pressing the Test push button, the Alarm and Fault LED indicators are immediately on. The buzzer is immediately on. Relays de-energize after 2 seconds of continuous pressing. Relays energize immediately

    Alarm on

    When pressing the Test push button, the Fault LED indicator is immediately on. Alarm LED indicator remains on. Buzzer remains on.

    The test will not affect relay 2, because it is already in alarm status. Fault alarm. When pressing the Test push button, the device does not react to the test at all.

    Altering settings

    Operating direction: High-level or low-level function (increasing or decreasing level). Operation delay: Two alternatives: 5 sec or 30 sec.

    Trigger level: Trigger point of an alarm in the sensor’s sensing element.

    Labkotec-SET-2000-110-VAC-Supply-Voltage

     

    Direction ModeLabkotec-SET-2000-110-VAC-Supply-Voltage

    Repair & Service

    The mains fuse (marked 315 mAT) can be changed to another glass tube fuse 5 x 20 mm / 315 mAT complying with N IEC 60127-2/3. Any other repair and service work on the device may be carried out only by a person who has received training in Ex-i devices and is authorized by the manufacturer.r

    Troubleshooting

    • Problem: The MAINS LED indicator is off.
    • Possible reason: Supply voltage is too low or the fuse is faulty. The transformer or MAINS LED indicator is faulty.
    • To do: Check if the two-pole mains switch is off. Check the fuse. Measure the voltage between the n poles + and -.
    • Problem: FAULT LED indicator is on.
    • Possible reason: Current in the sensor circuit too low (cable break) or too high (cable in short circuit).
    • To do: Measure the voltage separately between the poles 10 and 11. The voltages should be between 10,3….11,8 V. 3. If the voltages are correct, measure the sensor current.
    • Do as follows: Disconnect the sensor’s [+] wire from the sensor connector (pole 10). Measure short circuit current between [+] and [-] poles. Make a comparison to the values in the Table.

    Labkotec-SET-2000-110-VAC-Supply-Voltage

    Appendix 1 Technical data

    SET-1000
    Dimensions175 mm x 125 mm x 75 mm (L x H x D)
    EnclosureIP 65, material polycarbonate
    Cable glands4 pcs M16 for cable diameter 5-10 mm
    Operating environmentTemperature: -25 °C…+50 °C
    Max. elevation above sea level 2,000 m
    Relative humidity RH 100%
    Suitable for indoor and outdoor use (protected from direct rain)
    Supply voltage11-17 VDC
    Fuse 5 x 20 mm 315 mAT (EN IEC 60127-2/3)
    The device is not equipped with a mains switch
    Power consumptionMax. 1,2 VA.
    Typical (no alarms) 0,9 VA.
    SensorsOne Labkotec SET sensor
    Relay outputsTwo potential-free relay outputs
    250 V, 5 A, 100 VA
    Operational delay 5 sec or 30 sec. Relays de-energize at the trigger point. The operation mode is selectable for increasing or decreasing the level.
    Electrical safetyEN IEC 61010-1, Class II   , CAT II / III
    POLLUTION DEGREE 2
    Insulation level Sensor / Mains supply voltage375V (EN IEC 60079-11)
    EMC Emission ImmunityEN IEC 61000-6-3
    EN IEC 61000-6-2
    Ex-classification
    Special conditions (X)
    ATEX
    IECEx
    II (1) G    [Ex ia Ga] IIC
    (Ta = -25 C…+50 C) EESF 21 ATEX 022X
    IECEx EESF 21.0015X
    Year of manufacture
    See the serial number from the type plate
    xxx x xxxxx xx YY x
    where YY = year of manufacture
    (e.g. 19 = 2019)

     

    Appendix 2 Electrical parameters

    When installing the device, make sure that the electrical Labkotec SET-2000 110 VAC Supply Voltage values of the cable between SET-1000 and the sensor never exceed the maximum electrical parameters. The table below indicates the connecting values in explosion groups IIC and IIB. Uo = 14,7 V Io = 55 mA Po = 297 mW R = 404 Ω The characteristics of the output voltage is trapezoidal

    Labkotec-SET-2000-110-VAC-Supply-Voltage

    The maximum length of the sensor cable is determined by the resistance Labkotec SET-2000 110 VAC Supply Voltage (max. 75 Ω) and other electrical parameters (Co, Lo, and Lo/Ro) of the sensor circuit.

    . 81 Ω / km. – Inductance is approx. 3 μH / m. – Capacitance is approx. 70 nF/km Influence of resistance. The estimate for additional resistances in the circuit is 10 Ω.

    The max length of the cable is then (75 Ω – 10 Ω) / (81 Ω / km) = 800 m. The influence of inductance and capacitance of a oa800 m cable is: Influence of inductance. Total inductance is 0,8 km x 3 μH/m = 2,4 mH. The sum value of the cable and e.g,. SET/OS2 sensor [Li = 30 μH] is 2,43 mH. L/R ratio is thus 2,4 mH / (75 – 10) Ω = 37 μH/Ω, which is less than the maximum allowed value 116,5 μH/Ω. Influence of capacitance Cable capacitance is 0,8 km x 70 nF/km = 56 nF.

    Combined value of the cable and the e.g., SET/OS2 sensor [Ci = 3 nF] is 59 nF. When compared to the values in Table 2, we can summarize that the above values do not limit the use of this particular 800 m cable in explosion groups IIB or IIC.

    Customer Service

    • E-mail: labkonet@labkotec.fi
    • Tel: +358 29 006 6066.
    • Address: Myllyhaantie 6, FI-33960 Pirkkala, Finland

    © 2025 Labkotec.fi.

    FAQs

    What is the purpose of the Labkotec SET-2000?

    The SET-2000 is a control device designed to monitor separation systems (e.g., oil or grease separators) for the presence of liquid levels, such as oil, sludge, or water. It helps guarantee environmental safety and compliance by setting off alerts when the monitored medium surpasses a predetermined level.

    What power supply does this version of the SET-2000 use?

    The supply voltage for this particular model is 110 VAC. To prevent device failure or safety risks, be sure you are using the appropriate power source.

    Does the SET-2000 have any test or self-check features?

    Indeed. To make sure the indicator lights and relay outputs work properly, you can mimic an alarm condition using the device’s test button.

    Is the SET-2000 UL or CE certified?

    Indeed. For comprehensive certification information about the 110 VAC model for North American installations, refer to the handbook or get in touch with Labkotec.

    Can I connect the alarm output to a remote monitoring system?

    Indeed. Prior to wiring, review the output contact ratings found in the manual’s technical section.

    Labkotec SET-2000 110 VAC Supply Voltage installation guide Labkotec SET-2000 110 VAC Supply Voltage maintenance guide Labkotec SET-2000 110 VAC Supply Voltage manual Labkotec SET-2000 110 VAC Supply Voltage operation manual Labkotec SET-2000 110 VAC Supply Voltage user manual OnyxGuides Manual User Manual
    Share. Facebook Twitter Pinterest LinkedIn Tumblr Email
    Gabriel Smith

      Related Posts

      Ronix 8608 20V Cordless Electric Tool Instruction Manual

      November 4, 2025

      Ronix 4221 705W Electric Reciprocating Saw User Manual

      November 4, 2025

      Labnet S2056A VorTemp 56 Shaking Incubator User Manual

      November 4, 2025

      Labnet AccuPlate Laboratory Stirrer Hot Plate User Manual

      November 4, 2025

      Labnet H0600A ProBlot Hybridization Oven User Manual

      November 4, 2025

      Labnet I5110A Mini Microbiology Incubator User Manual

      November 4, 2025
      Add A Comment

      Comments are closed.

      Don't Miss
      Ronix

      Ronix 8608 20V Cordless Electric Tool Instruction Manual

      November 4, 2025

      Ronix 8608 20V Cordless Electric Tool Safety Instructions Keep the work area clean. Cluttered areas…

      Ronix 4221 705W Electric Reciprocating Saw User Manual

      November 4, 2025

      Labnet S2056A VorTemp 56 Shaking Incubator User Manual

      November 4, 2025

      Labnet AccuPlate Laboratory Stirrer Hot Plate User Manual

      November 4, 2025
      Stay In Touch
      • Facebook
      • Twitter
      • Pinterest
      • Instagram
      • YouTube
      • Vimeo

      Subscribe to Updates

      Subscribe to Updates

      Follow Us

      Facebook Twitter Youtube

      © 2025 All rights reserved Onyx Guide.