Labnet C1001 MPS 1001 Mini Plate Spinner

Introduction
The Labnet MPS-1001 Mini Plate Spinner is designed for quick spins of samples in PCR plates. Primarily designed for use with 96- or 384-well plates and small volumes, it will accommodate skirted, non-skirted, and semi-skirted styles. With the use of the adapter, you can spin a half plate, 8 or 12 x 0.2 μL strips, and individual PCR tubes. No decontamination is required for the equipment. Before using any cleaning or decontamination methods except those recommended by the manufacturer, check with the manufacturer that the proposed method will not damage the equipment.
Safety Information
Never use the centrifuge in any manner not specified in these instructions.
- Never move the centrifuge while the rotor is spinning.
- Never lean on or leave the centrifuge during operation.
- Never use solvents or flammables near this or other electrical equipment.
- Never centrifuge flammable, explosive, or corrosive materials.
- Never centrifuge hazardous materials outside of a hood or proper containment facility.
- Never centrifuge in a strong ultraviolet environment.
- Never continue to operate a centrifuge that is vibrating excessively.
- Always place the centrifuge on a level, sturdy surface.
- Always load plates so that the sealed side faces the center of the rotor and the bottom of the plate faces out.
- Always locate the centrifuge within easy access to an electrical outlet.
Installation
Plug the power cord into the back of the unit, and connect the other end to the power socket. To turn on the unit, set the power switch to the On (I) position.
Operation

To avoid spillage of samples, only plates that have been heated or ortape-sealedd should be placed in the centrifuge. Plates are loaded into the rotor in a vertical position.
- Lift the clear centrifuge lid to gain access to the rotor. Carefully, slide the first plate into position so that the sealed side faces the center of the rotor and the bottom of the plate faces out.
- Load the second plate in the same manner as the first, making sure that it is all the way down in the rotor slot.
- The time setting range is from 0 to 99 minutes. Press the time increase/decrease button to adjust the time parameter. Select 0 to cancel the timing function, and press the START/STOP button to save the data.
- To begin a run, close the lid of the centrifuge. Press the START/STOP button to run. Allow the unit to run for 15-20 seconds. This is sufficient time to spin the samples down to the bottoms of the wells.
- To stop, press the START/STOP button, and the rotor will decelerate. Once the rotor has stopped, press the Open Door button to open the centrifuge lid and take out the sample plate(s).
Cleaning
Be sure to disconnect the power cord before cleaning. After the rotor is stopped, clean the surface of the instrument with a dry cloth and 75% alcohol. Make sure that all parts of the instrument are completely dry before operating. Do not immerse the unit in liquid.
Specifications
| Capacity | 2 x 96-/384-well PCR plates |
| Dimensions | 7.5 x 8.6 x 7.3 in. (190 x 220 x 186 mm) |
| Weight | 5.7 lbs. (2.6 kg) |
| Power supply | 100 to 240V~, 50/60 Hz |
| Maximum speed | 2,500 rpm ±10% |
| Maximum RCF | 500 x g |
| Temperature range | 5°C to 40°C (up to 80% RH at 31°C, decreasing linearly to 50% RH at 40°C) |
| Electrical | 100V, 50/60Hz, 0.6A 120V, 50/60Hz, 0.6A 230V, 50/60Hz, 0.3A |
| Protection class | IP21 |
- Indoor use
- Altitude up to 2,000 meters
- Pollution Degree 2
Limited Warranty
Corning Incorporated (Corning) warrants that this product will be free from defects in material and workmanship for a period of one (1) year from the date Labnet C1001 MPS 1001 Mini Plate Spinner of purchase. Corning’s sole obligation shall be to repair or replace, at its option, any product or part thereof that proves defective in material or workmanship within the warranty period, provided the purchaser notifies Corning of any such defect. Corning is not liable for any incidental or consequential damages, commercial loss, or any other damages from the use of this product.
Corning’s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions. Corning may elect for on-site service for larger equipment. Some states do not alllimitationsion on the length of implied warranties or the exclusion or limitation of incidental or consequential damages. This warranty gives you specific legal rights. You may have other rights that vary from state to state. No individual may accept for, or on behalf of Corning, any other obligation of liability, or extend the period of this warranty
Warranty/Disclaimer
Unless otherwise specified, all products are for research use or general laboratory use only.* Not intended for use in diagnostic or therapeutic procedures. Not for use in humans. Corning Life Sciences makes no claims regarding the performance of these products for clinical or diagnostic applications. *For a listing of US medical devices, regulatory classifications, or specific information on claims, visit www.corning.com/resources.
Equipment Disposal
According to Directive 2012/19/EU of the European Parliament and of the Council of 4 July 2012 on waste electrical and electronic equipment (WEEE), this product is marked with the crossed-out wheeled bin and must not be disposed of with domestic waste. Consequently, the buyer shall follow the instructions for reuse and recycling of waste electronic and electrical equipment (WEEE) provided with the products and available at www.corning.com/weee.
Customer Service
- Website: www.labnetlink.com.
- Visit: www.corning.com/lifesciencs
- Visit: www.corning.com/resources
- USA/Canada: 1.800.492.1110
- Others: +1.978.442.2200
- North America: 978.442.2200
- Australia/New Zealand: 61 427286832
FAQs
How do I adjust the spinning speed?
The apparatus has a set speed of about 1000 RPM, which is ideal for efficient plate spinning without disturbing the sample.
Is the Mini Plate Spinner suitable for continuous operation?
No, the device is made for brief spin cycles, usually lasting between a few seconds and a minute. It is not advised to spin for extended periods of time.
How do I start and stop the spin cycle?
To start spinning, press the start button. After a predetermined amount of time or when the stop button is manually hit, the device will automatically cease.
Does the unit have a timer?
Indeed, a timer with preset spin durations ranging from 1 to 99 seconds is provided.
How do I clean and maintain the Mini Plate Spinner?
Use a gentle cloth and a light detergent to clean the rotor and exterior. Prevent water from getting into the electronic or motor components.