Labnet P2002 FastPette Pro Pipet Controller
Introduction
The pipette controller is a device intended for general laboratory use only, for pipetting liquids with the use of measuring pipettes. It can work with all types of glass or plastic pipets in the volume range from 0.5 mL to 100 mL. Two dispensing modes permit selection of dispensing intensity depending on the user’s needs.
Safety Information
Use of this product in any manner not specified by the manufacturer may impair the safety protection provided by the equipment and may result in physical damage, personal injury, and/or errors in pipetting of liquids. Please read all operating instructions in this manual before using this equipment.
This symbol indicates a potential risk and alerts you to proceed with caution.
- Using the device inconsistently with the operating instructions may result in damaging the device.
- The pipet controller may work in a temperature range from +10°C to +35°C.
- The pipette controller is suitable for general laboratory use only.
Switching On
The pipet controller is switched on by pressing any of the buttons (Figure 1A, B, C, D).
The display will show the selected aspiration mode, dispense mode, and the battery level indicator. 
- The pipet controller switches off automatically if not used for 5 minutes.
- The mains voltage shall conform to the specification on the charger.
Aspirating and Dispensing Liquids
Attaching a pipette
Before attaching a pipette, check whether the pipette is not damaged, has no dents or sharp edges in the gripping part. Check whether the gripping part is dry. 
A pipette that has been correctly attached and sealed in the holder should not tilt to the sides. After attaching a pipette, hold the pipette controller in a vertical position.
Filling the pipette
Before aspirating is started, set the speed by pressing the SPEED switch (Figure 1C) until the display shows the correct speed (Figure 1E).
- HIGH speed: fast aspirating (Figure 2.1)
- LOW speed: slow aspirating (Figure 2.2)
Setting the volume
After the pipette is filled, dry the outside surface with absorbent paper that does not leave impurities. Then set the required liquid volume precisely. Pressing the dispense button gently (Figure 3.3), dispense the excessive liquid from the pipette until the meniscus of the liquid aligns exactly with the required volume mark on the pipette.
Emptying the pipette
Holding the vessel in an inclined position, place the pipet end in contact with the vessel wall and press the dispense button gently (Figure 3.3).
Troubleshooting
If during your work the pipet controller is not functioning correctly, check the cause and correct the fault.
| Problem Possible Cause | Action | |
| The pipette falls out. The pipette holder (Figure 1G) (the holding force of it is dirty or wet. | Take out the pipette holder, and clean, wash, and dry it | |
| The pipette is too small, or is tooto the side too The pipette holder is damaged. Much. | Replace the pipette holder with a new one. | |
| The pump is working. The filter (Figure 1H) is dirty. But the pipet controller | Take out the pipette holder, take out the filter. If it is dirty, replace it with a new one. | |
| Does not draw liquid or draws liquid very slowly. | The pipet holder and/or the connector gasket (Figure 1J) is damaged. | Replace the mechanically damaged elements with new ones. |
Replacing the Filter
The direct reason may be a dirty filter after a long period of use. To replace the filter:
- Remove the pipette.
- Unscrew the nosepiece (Figure 4.1).
- Remove the membrane filter (Figure 4.1) and the pipette holder (Figure 4.2).
- Rinse the holder using a wash bottle (Figure 4.3).
- Blow liquid out of the holder and set it aside until it is completely dry.
- Install a new membrane filter (Figure 4.4), and assemble the device in reverse order.

Charging the Batteries
Using chargers other than the original one may damage the battery. The pipet controller requires 3 NiMH type AAA batteries (included). 
Charging
- Charging temperature: 10°C to 55°C. Batteries’ charging is indicated by the successive lighting of the “bars”.
- Full charging time: 7 to 8 hours. The service life of the batteries: approx. 1,000 charging cycles, if used correctly.
Charging on a charging stand

- Put the pipet controller on the stand.
- Plug the charger into the stand outlet.
- Connect the charger to the adapter and to the main power.
- If the pipet controller starts to indicate a low battery level during work, connect it to the charger to continue working.
- Do not leave the pipet controller discharged for a long period of time.
Maintenance
Ultraviolet (UV) sterilization
The recommended distance from the radiation source to the exposed element should be not less than 50 cm. Prolonged and very intense UV exposure can cause discoloration of pipet controller parts, without affecting its performance.
Storage
The pipet controller should be stored in a dry place. The allowable storage temperature is -20°C to +50°C. During breaks in the work, the pipet controller can be stored on the wall hanger or charging stand.
Components
- Universal charger with a set of adapters
- PTFE filter 0.2 µm
- Batteries (3 AAA)
- Charging stand
- Quick Start Guide
- QC certificate
Ordering Information
The Labnet FastPette Pro Pipet Controller (Cat. No. P2002) comes with a universal charger and a set of adapters in different versions (EU, US, UK, and AU). Choose your country’s adapter and connect to the housing. To remove or change the adapter, simply press the PUSH button in the direction of the arrow, hold the button down, and move the adapter in the direction of the arrow.
Spare Parts
| Item in Figure 1 | Description | Cat. No. | Qty/Pk |
| F | Nosepiece | SP29052 | 1 |
| G | Silicone pipette holder | SP29054 | 1 |
| H | PTFE filter 0.2 µm | SP9143 | 5 |
| PTFE filter 0.45 µm | SP9144 | 5 | |
| L | Battery set, 3xAAA, NiMH | SP29066 | 3 |
| M | Charging stand | SP29055 | 1 |
| N | Universal charger, 9V with a set of adapters (EU, US, UK, AU) | SP29100 | 1 |
| K | Battery cover | SP29062 | 1 |
| P | Wall mount | SP9029 | 1 |
Limited Warranty
Corning Incorporated (Corning) warrants that this product will Labnet P2002 FastPette Pro Pipet Controller be free from defects in material and workmanship for a period of one (1) year from the date of purchase. Corning’s sole obligation shall be to repair or replace, at its option, any product or part thereof that proves defective in material or workmanship within the warranty period, provided the purchaser notifies Labnet P2002 FastPette Pro Pipet ControllerCorning of any such defect. Corning is not liable for any incidental or consequential damages, commercial loss, or any other damages from the use of this product.
Corning’s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions. Corning may elect for on-site service for larger equipment. Some states do not alllimitationsion on the length Labnet P2002 FastPette Pro Pipet Controller of implied warranties or the exclusion or limitation of incidental or consequential damages. This warranty gives you specific legal rights. You may have other rights that vary from state to state. No individual may accept for, or on behalf of Corning, any other obligation of liability, or extend the period of this warranty
Warranty/Disclaimer
Unless otherwise specified, all products are for research use or general laboratory use only.* Not intended for use in diagnostic or therapeutic procedures. Not for use in humans. Corning Life Sciences makes no claims regarding the performance of these products for clinical or diagnostic applications. *For a listing of US medical devices, regulatory classifications, or specific information on claims, visit www.corning.com/resources.
Equipment Disposal
Consequently, the buyer shall follow the instructions for reuse and recycling of waste electronic and electrical equipment (WEEE) provided with the products and available at www.corning.com/weee.
Customer Service
- Website: www.labnetlink.com.
- Visit: www.corning.com/lifesciencs
- Visit: www.corning.com/resources
- USA/Canada: 1.800.492.1110
- Others: +1.978.442.2200
- North America: 978.442.2200
- Australia/New Zealand: 61 427286832
FAQs
What types of pipettes are compatible with the FastPette Pro P2002?
Single-channel glass and plastic serological pipettes with capacities ranging from 0.1 mL to 100 mL are compatible with the P2002.
How is the airflow controlled on the FastPette Pro?
The apparatus has simple push-button controls for dispensing and aspirating, enabling accurate control of airflow.
What powers the FastPette Pro P2002?
A rechargeable lithium-ion battery powers the device, allowing for up to eight hours of continuous operation between charges.
How do I recharge the battery?
Just plug the supplied charger into the charging port on the device.
Is the FastPette Pro easy to clean and maintain?
Yes, you can use a moist towel to wipe the outside. It is possible to separate and clean the pipette tip and holder.
Does the P2002 include any safety features?
Indeed, it contains an integrated overpressure prevention system to guard against pipette damage and lessen user fatigue.