Labnet P2000 FastPette V2 Pipette Controller
Introduction
The pipette controller is a device intended for general laboratory use only, for pipetting liquids with the use of measuring pipettes. It can work with all types of glass or plastic pipets in the volume range from 0.5 mL to 100 mL. Two dispensing modes permit selection of dispensing intensity depending on the user’s needs (Figure 1D). The FastPette V-2 features a two-speed control system, which enables very fast dispensing of large volumes and precise measuring of small volumes. Figure 1 shows the external parts of the pipet controller with a description of the materials used.
Safety Information
Risk of damage to the device or errors in the pipetting of liquids. Before starting the work with the pipet controller, every user should read these operating instructions carefully. Using the device inconsistently with the operating instructions may result in damaging the device.
Product Description

- Aspiration button – PP
- Dispense button – PP
- Suction speed switch – PP
- Dispense mode switch – PP
- Indicator
- Nose piece – PP
- Pipet holder – SI
- Membrane filter – PP/PTFE
- Connector gasket – SI
- Bench stand
- Charger 9V: EU, US, UK, AU
- INPUT: 100-240V, 50/60Hz, 0.3A
- OUTPUT: DC 9V, 230mA
- Wall mount – PP
Switching On
The pipet controller is switched on by pressing the trigger buttons (Figure 1A, B, C, D). Charge the batteries before the first use. Full charging cycle takes at least 11 hours.
- The pipet controller may be charged only with the original charger.
- The mains voltage shall conform to the specification on the charger.
- Charging shall be done in accordance with section 8 of the instruction manual.
Aspirating and Dispensing Liquids
Attaching a pipette
Check whether the gripping part is dry. Do not apply excessive force so as not to damage thin pipets and avoid the risk of injury. After attaching a pipette, hold the pipette controller in a vertical position.
Filling the pipette
Before aspirating is started, set the speed by using the SPEED switch.
- HIGH speed – fast aspirating,
- LOW speed – slow aspirating.
Setting the volume
After the pipette is filled, dry the outside surface with absorbent paper that does not
leave impurities. Then set the required liquid volume precisely. Pressing the dispense
button gently (Figure 3.3), dispense the excessive liquid from the pipette until the
meniscus of the liquid aligns exactly with the required volume mark on the pipette.
Emptying the pipette
Holding the vessel in an inclined position, place the pipet end in contact with the vessel wall and press the dispense button gently (Figure 3.3).
- Gravity mode – dispensing is effected in gravity mode, which means that the liquid flows out of the pipette by its own weight.
- Blow out mode – dispensing is effected in gravity mode; however, when the dispense button is pressed to the middle position, the pump is started, and fast emptying of the pipette with a blow out is effected.
Replacing the Filter
The direct reason may be a dirty filter after a long period of use. To replace the filter:
- Remove the pipette.
- Unscrew the nose piece (Figure 4.1).
- Remove the membrane filter (Figure 4.1) and the pipette holder (Figure 4.2).
- Rinse the holder using a wash bottle (Figure 4.3).
- Blow liquid out of the holder and set it aside until it is completely dry.
- Install a new membrane filter (Figure 4.4) and assemble the device in reverse order.
Charging the Batteries
Using chargers other than the original one may damage the battery.
- Charging temperature: 10°C to 55°C.
- Full charging time: 11 to 14 hours..
- If the pipet controller starts to indicate a low battery level during work, connect it to the charger to continue working.
- Do not leave the pipet controller discharged for a long period of time.
Maintenance
Ultraviolet (UV) sterilization
The recommended distance from the radiation source to the exposed element should be not less than 50 cm. Prolonged and very intense UV exposure can cause de-coloration of pipet controller parts, without affecting its performance.
Storage
The pipet controller should be stored in a dry place. The allowable storage temperature: -20°C to +50°C. During breaks in the work, the pipet controller can be stored on the wall hanger or bench stand.
Components
- Universal charger with a set of adapters
- PTFE filter 0.2 µm
- Instruction manual
- Bench stand
- QC certificate
Troubleshooting
If during your work the pipet controller is not functioning Labnet P2000 FastPette V2 Pipette Controller correctly, check the cause and correct the fault.
| Problem Possible Cause | Action | |
| The pipette falls out. The pipette holder (Figure 1G) (the holding force of which is dirty or wet. | Take out the pipette holder, and clean, wash, and dry it. | |
| The pipette is too small, or it tilts to the side too. | Replace the pipette holder with a new one. | |
| The pump is working. The filter (Figure 1H) is dirty. But the pipet controller | Take out the pipette holder, take out the filter; if it is dirty, replace it with a new one. | |
| Aid does not draw liquid or draws liquid very slowly. | The pipet holder and/or the connector gasket (Figure 1J) is damaged. | Replace the mechanically damaged elements with new ones. |
Ordering Information
The Labnet FastPette V2 Pipet Controller comes with a universal charger and a set of adapters in different versions: EU, US, UK, and AU. Choose your country’s adapter and connect to the housing.
To remove or change the adapter, simply press the “PUSH” button in the direction of the arrow, hold the button down, and remove the adapter in the direction of the arrow.
Spare Parts
| Item in Figure 1 | Description | Cat. No. | Qty/Pk |
| F | Nose piece | SP9022 | 1 |
| G | Silicone pipette holder | SP29054 | 1 |
| H | PTFE filter 0.2 µm | SP9143 | 5 |
| PTFE filter 0.45 µm | SP9144 | 5 | |
| M | Bench stand | SP19030 | 1 |
| N | Universal charger, 9V with a set of adapters: EU, US, UK, AU | SP29100 | 1 |
| P | Wall mount | SP9029 | 1 |
Limited Warranty
Corning Incorporated (Corning) warrants that this product Labnet P2000 FastPette V2 Pipette Controller will be free from defects in material and workmanship for a period of one (1) year from the date of purchase. Corning’s sole obligation shall be to repair or replace, at its option, any product or part thereof that proves defective in material or workmanship within the warranty period, provided the purchaser notifies Corning of any such defect. Corning is not liable for any incidental or consequential damages, commercial loss, or any other damages from the use of this product.
Corning’s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions. Corning may elect for on-site service for larger equipment. Some states do not alllimitationsion on the length of implied warranties or the exclusion or limitation of incidental or consequential damages. This warranty gives you specific legal rights. You may have other rights that vary from state to state. No individual may accept for, or on behalf of Corning, any other obligation of liability, or extend the period of this warranty.
Warranty/Disclaimer
Unless otherwise specified, all products are for research use or general laboratory use only.* Not intended for use in diagnostic or therapeutic procedures. Not Labnet P2000 FastPette V2 Pipette Controller for use in humans. Corning Life Sciences makes no claims regarding the performance of these products for clinical or diagnostic applications. *For a listing of US medical devices, regulatory classifications, or specific information on claims, visit www.corning.com/resources.
Equipment Disposal
Consequently, the buyer shall follow the instructions for reuse and recycling of waste electronic and electrical equipment (WEEE) provided with the products and available at www.corning.com/weee.
Customer Service
- Website: www.labnetlink.com.
- Visit: www.corning.com/lifesciencs
- Visit: www.corning.com/resources
- USA/Canada: 1.800.492.1110
- Others: +1.978.442.2200
- North America: 978.442.2200
- Australia/New Zealand: 61 427286832
FAQs
What is the Labnet FastPette V2 used for?
The FastPette V2 is a motorised pipette controller that is perfect for cell culture, media preparation, and routine laboratory work since it makes it easier and faster to aspirate and dispense liquids using serological pipettes ranging in capacity from 1 to 100 mL.
What types of pipettes are compatible with the FastPette V2?
Serological pipettes with standard thin necks that range in capacity from 1 mL to 100 m,L made of glass or plastic are compatible with the device.
How is the aspirate and dispense speed controlled?
The FastPette V2 has an adjustable speed control dial and a pressure-sensitive trigger. Full pressure accelerates the pace for rapid transfers, while light pressure permits slower speeds for accuracy.
Is the unit battery-operated or rechargeable?
Indeed, the FastPette V2 features a rechargeable lithium-ion battery that, when fully charged, enables up to 8 hours of continuous cordless use.
How do I know when the battery is low or fully charged?
The device displays the current battery levels via an LED battery status indicator. When the battery is low or charging, it will flash or change colour.
Can I adjust between gravity and motor-driven dispensing?
Indeed. For thick or residual samples, the FastPette V2 offers a motor-powered blow-out option for full liquid release and a gravity dispense mode for gentle dispensing.