Congratulations on purchasing your Labnet ProBlot™ Rocker. This high-capacity bench rocker is a basic bench-rocking platform that has been designed for long life and trouble-free service. The bench rocker speed is adjustable to provide the gentle motion required for staining fragile gels as well as the more vigorous action necessary for washing blots.
Safety Information
Do not use the rocker near water. Ensure that water doesn’t get on the device, particularly during cleaning.
- If the unit is used in a manner not specified by the manual, the degree of protection may be impaired.
- The plug should only be used in a wall outlet with protective earthing.
- The unit should be used with a power supply cord of at least 18 AWG.
- The equipment will not restart if mains power is interrupted.
- The equipment will restart in cases of mechanical interruption.
- Do not expose the rocker to aggressive vapors (e.g., organic solvents).
- Do not insert any objects into the rocker (such as a screwdriver or metal rods).
- Never operate the rocker without the platform securely attached.
- Hazardous samples should be shaken only in appropriate containment vessels
Unpacking
The Labnet ProBlot Rocker includes the following components:
- S2025-B ProBlot Rocker with single platform
- S2025-D-B ProBlot Rocker with double platform
- S2025-XL-B ProBlot Rocker with a large single platform
- S2025-XLD-B ProBlot Rocker with a large double platform
- Instruction manual
Installation
Device Placement
When selecting the right place for the device, consider the following:
- Place the device on a smooth, horizontal, and stable surface.
- Leave enough space around the device for normal air circulation, min. 10 cm.
- Leave enough space around the device for easy access and maintenance.
- Avoid locations subject to shocks and vibrations.
- Avoid locations subject to frequent power fluctuations or power losses, as this may affect the shaker
performance..
Attaching the Power Cord
First, connect the main power cord to the shaker, then connect the plug into a grounded wall socket.
Avoid sockets where powerful electric motors, refrigerators, and similar devices are also connected.
- Be careful when you plug the cord into the grounded wall socket.
- Do not touch the plug with wet hands.
- Do not pull the plug by the cord.
Installation of Double Stacking Platform
- Set posts (A) through the four holes at the corners of the bottom platform.
- Add a washer (B) and nut (C) to each post.
- Fit the upper platform (D) onto the top of the four posts through the corner holes.
- Screw on the four metal knurl nuts (E) to secure the upper platform.

Control Panel

- POWER switch – to turn the rocker ON (power switch illuminates) or OFF.
- SPEED controller – to set the shaker speed.
- TIME controller – to set the timer.
- RUN – green light illuminates when the shaker is operating.
- START/STOP button – to START or STOP the shaking operation.
- HOLD function – the yellow light illuminates and the shaker operates continuously
Basic Operation
- Switch on POWER.
- Set the SPEED to the desired value.
- Set the TIME to the desired value or to HOLD (the shaker will operate continuously and the yellow light will illuminate).
- Press the START/STOP button to start rocking.
- Press the START/STOP button to stop rocking when in the HOLD position. The device will automatically stop
after the set time when using the timer.
Maintenance and Cleaning
- No routine maintenance or lubrication is required.
- It is recommended that the rocker be kept clean to ensure long, trouble-free operation.
Cleaning
Prior to any cleaning or maintenance of the device, unplug the power cord from the wall socket.
The outside housing can be cleaned with a soft, damp cloth when needed. Do not use any aggressive or
abrasive cleaners (acetone, nitro, polish, etc.) This may permanently damage the rocker finish.
Specifications
| Electrical requirements | Rocker 115V: 120V± 10% – 50/60Hz, 0.2A Rocker 230V: 230V± 10% – 50/60Hz, 0.1A |
| Fuse | 2 x 1A (120V) 2 x 0.5A (230V) |
| Speed settings | 5 – 120 tilts/minute |
| Motion | Rocking |
| Tilt range | Fixed 7° |
| Timer | from 0 to 2 hrs. or continuous |
| Maximum load | 11 lbs. (5 kg) |
| Dimensions | Rocker 25: 11.4 x 12.6 x 5.7 in. (29 x 32 x 14.5 cm) Rocker 25 XL: 12.8 x 14.2 5.7 in. (32.5 x 36 x 14.5 cm) |
| Platform dimensions | 11.4 x 8 in. (28.9 x 20.3 cm) |
| Weight | 11.7 lbs. (5.6 kg) |
| Operating conditions | from 4°C to 65°C, up to 85 RH, non-condensing |
Troubleshooting
Should you have a question about the Labnet ProBlot™ Rocker 25 and 25XL or require service for the unit,
Contact Corning Customer Service at 800.492.1110 or 978.442.2200. Do not send the unit for service
without first calling to obtain a repair authorization (RA) number. The unit should be properly packed to avoid damage. Any damage resulting from improper packaging shall be the responsibility of the user.

Limited Warranty
Corning Incorporated (Corning) warrants that this Labnet S2025-B ProBlot Rocker 25 and 25XL product will be free from defects in material and workmanship for a period of one (1) year from the date of purchase. Corning’s sole obligation shall be to repair or replace, at its option, any product or part thereof that proves defective in material or workmanship within the warranty period, provided the purchaser notifies Corning of any such defect. Corning is not liable for any incidental or consequential damages, commercial loss, or any other damages from the use of this product.
Corning’s Customer Service team will help Labnet S2025-B ProBlot Rocker 25 and 25XL arrange local service where available or coordinate a return authorization number and shipping instructions. Corning may elect for on-site service for larger equipment. Some states do not alllimitationsion on the length of implied warranties or the exclusion or limitation of incidental or consequential damages. This warranty gives you specific legal rights. You may have other rights that vary from state to state. No individual may accept for, or on behalf of Corning, any other obligation of liability, or extend the period of this warranty.
Warranty/Disclaimer
Unless otherwise specified, all products are for research use or general laboratory use only.* Not intended for use in diagnostic or therapeutic procedures. Not for use in humans. Corning Life Sciences makes no claims regarding the performance of these products for clinical or diagnostic applications. *For a listing of US medical devices, regulatory classifications, or specific information on claims, visit www.corning.com/resources.
Equipment Disposal
According to Directive 2012/19/EU of the European Parliament and of the Council of 4 July 2012 on waste electrical and electronic equipment (WEEE), this product is marked with the crossed-out wheeled bin and must not be disposed of with domestic waste. Consequently, the buyer shall follow the instructions for reuse and recycling of waste electronic and electrical equipment (WEEE) provided with the products and available at www.corning.com/weee.
Customer Service
- Website: www.labnetlink.com.
- Visit: www.corning.com/lifesciencs
- Visit: www.corning.com/resources
- USA/Canada: 1.800.492.1110
- Others: +1.978.442.2200
- North America: 978.442.2200
- Australia/New Zealand: 61 427286832
FAQs
What are the Labnet ProBlot Rocker 25 and 25XL used for?
These rockers are ideal for Western blotting, gel staining/destaining, membrane washing, and hybridisation procedures because they are made to mix materials gently and consistently.
What is the difference between the 25 and 25XL models?
While the 25XL has an extra-large platform that can accommodate more or larger vessels, making it perfect for high-throughput workflows, the ProBlot 25 has a lower platform size that is appropriate for typical lab use.
What is the rocking motion like?
With a preset tilt angle of ±7°, all versions provide a see-saw rocking motion that guarantees even liquid distribution across samples without splashing.
Is the speed adjustable?
Indeed. The variable speed control on both models, which ranges from 5 to 120 RPM, enables users to adjust the agitation intensity based on the application.
Can I set a timer on the rocker?
There is no built-in timer in the ordinary variants. If timed runs are required, users can utilise an external timer to operate the devices.
Is the ProBlot Rocker safe for use in cold rooms or incubators?
Indeed. The 25 and 25XL are made to work in temperature-controlled settings, such as cold rooms and non-CO₂ incubators, between +4°Cand 65 65°C.
What types of containers can I use on the platform?
Petri dishes, blot boxes, flasks, culture plates, and bags can all be stored on the platforms. Included are non-slip mats to assist in keeping containers secure when rocking.
Can I replace or customize the platform?